Description
Was digging around for the definition of trapTmax. The tutorial seems to have a nice illustration with the processor
"trapTmax": {
"function": "argmax",
"module": "numpy",
"args": ["wf_trap", 1, "trapTmax"],
"kwargs": {"signature": "(n),()->()", "types": ["fi->i"]},
"unit": "ns",
}that indicates that it should a time (note that it has units of ns). The attached image in the documentation also seems to show a vertical blue line that presumably corresponding to a time.
HOWEVER, digging through the config files in legend-dataflow-config (which now have short descriptions of the parameters), it is clear that trapTmax is the max of a fixed length long trap filter and should correspond to an uncalibrated energy. Here is the relevant section from a config file
"trapTmax": {
"description": "max of fixed length long trap filter",
"function": "amax",
"module": "numpy",
"args": [
"wf_trap[round(2*db.ttrap.rise+db.ttrap.flat, wf_pz.period):-round((2*db.ttrap.rise+db.ttrap.flat), wf_pz.period)]",
1,
"trapTmax"
],
"kwargs": { "signature": "(n),()->()", "types": ["fi->f"] },
"defaults": { "db.ttrap.rise": "10*us", "db.ttrap.flat": "3.008*us" },
"unit": "ADC"
}Could we make the tutorial definition consistent with the processor being used for production?