Skip to content

GNARL battery-test with Loop #2

@cfaagaard

Description

@cfaagaard

hi,

I have made tests with the following setup:

IMG_20191101_200120

Test:

  • 10 hours charging of battery on GNARL-hardware
  • Running Loop connected to Medtronic Pump and GNARL
  • Did some bolus, suspend pump, overrides, changed settings
  • Used Nightscout to collect data via Loop (version 0.12.5)

Result:
I did the test two times and both tests showed a battery-lifetime just around 18 hours.
This is compared to the Riley link which have 24-30 hours battey-lifetime and 2 hours recharging.

Observations/suggestions:

  • Wifi: I can see that wifi-module is included in the code. In the GNARL/Loop scenario wifi is not needed. If the wifi-hardware is turned on, turning this off will increase batterylifetime.

  • No Bluetooth-connected indication on the GNARL (like Rileys green light).

  • GNARL sleep-modes: Maybe we should look more into this. But I do not think it possible to wake it up from a bluetooth signal.

  • the battery charging time is an issue (8-10 hours).