Notifications
You must be signed in to change notification settings - Fork 0
/
Copy pathCheckError.m
129 lines (129 loc) · 7.07 KB
/
CheckError.m
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
function [message] = CheckError(error_code)
% Module Related Errors
% E 0101 Error in the low wavelength monitor current.
% E 0102 Error in the high wavelength monitor current.
% E 0103 Error in the low wavelength laser diode current.
% E 0104 Error in the high wavelength laser diode current.
% E 1xxx LIGHT Error while zeroing. Digits marked x indicate the phase, loop
% counter value, and hardware.
% E 3000 NO HEAD No head attached to the optical head interface module.
% E 3100 SETTLING Module not yet adjusted to temperature.
% E 3200 Checksum error in optical head data.
% E 3300 No memory to store the zero data (memory allocation failure).
% E 3400 NO HEAD Optical head removed during the zero operation.
% E 4001 MODULE Reading EEPROM failed.
% E 4002 MODULE EEPROM checksum error.
% E 4004 MODULE No acknowledge from EEPROM.
% E 4005 MODULE reading EPROM failed.
% E 4006 KEY JAM reading EPROM failed.
%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
% Specific Error Identifiers
% E 4010 Input value exceeds limits
% E 4020 Input value below lower limit
% E 4030 Registers of the real time clock cannot be read from or written to.
%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
% Store and Recall Errors
% E 4101 CH EMPTY Store from or recall to an empty slot.
% E 4102 MISMATCH Attempt to recall source data into a sensor module, or
% sensor data into a source module.
% E 4103 NO DATA Attempt to recall data from a location where no data has
% been stored, or where the data is unintelligible.
% E 4105 MEM FAIL Memory checksum failed.
%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
% Plot, Print, Show, and Manual Logging Errors
% E 4201 NO DATA There is no data for the application. Another cause could be
% that there is not enough data for the results statistics to be calculated.
% E 4202 INVALID The data is invalid.
% E 4203 TLKONLY? Plot or print is not possible because the HP-IB is not set to
% talk only mode.
%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%
% Loss Errors
% E 4301 CONFIG The Loss application cannot run, because it
% needs both a source and a sensor module.
% E 4302 CONFIG The
% Loss application cannot run, because it needs one of the
% modules is already being used.
if isempty(error_code)
message = 'no errores';
else
switch lower(error_code)
% Module Related Errors
case 'e 0101'
error('Error in the low wavelength monitor current.');
case 'e 0102'
error('Error in the high wavelength monitor current.');
case 'e 0103'
error('Error in the low wavelength laser diode current.');
case 'e 0104'
error('Error in the high wavelength laser diode current.');
case 'e 1xxx'
error(['LIGHT Error while zeroing.'...
' Digits marked x indicate the phase,'...
' loop counter value, and hardware.']);
case 'e 3000'
error(['NO HEAD No head attached to'...
' the optical head interface module. ']);
case 'e 3100'
error(' SETTLING Module not yet adjusted to temperature. ');
case 'e 3200'
error('Checksum error in optical head data.');
case 'e 3300'
error(['No memory to store the zero data'...
' (memory allocation failure).']);
case 'e 3400'
error(['NO HEAD Optical head removed'...
' during the zero operation.']);
case 'e 4001'
error('MODULE Reading EEPROM failed.');
case 'e 4002'
error('MODULE EEPROM checksum error. ');
case 'e 4004'
error('MODULE No acknowledge from EEPROM.');
case 'e 4005'
error('MODULE reading EPROM failed.');
case 'e 4006'
error('KEY JAM reading EPROM failed');
% Specific Error Identifiers
case 'e 4010'
error('Input value exceeds limits');
case 'e 4020'
error('Input value below lower limit');
case 'e 4030'
error(['Registers of the real time clock'...
' cannot be read from or written to.']);
%Store and Recall Errors
case 'e 4101'
error('CH EMPTY Store from or recall to an empty slot.');
case 'e 4102'
error(['MISMATCH Attempt to recall source data'...
' into a sensor module, or '...
'sensor data into a source module.']);
case 'e 4103'
eerror(['NO DATA Attempt to recall data'...
' from a location where no data has'...
' been stored, or where the data is unintelligible.']);
case 'e 4105'
error('MEM FAIL Memory checksum failed.');
% Plot, Print, Show, and Manual Logging Errors
case 'e 4201'
error(['NO DATA There is no data for the application.'...
' Another cause could be that there is not enough'...
' data for the results statistics to be calculated']);
case 'e 4202'
eerror('INVALID The data is invalid.');
case 'e 4203'
error(['E 4203 TLKONLY? Plot or print is not possible'...
' because the HP-IB is not set to talk only mode.']);
% Loss Errors
case 'e 4301'
eerror(['CONFIG The Loss application cannot run,'...
' because it needs both a source'...
' and a sensor module.']);
case 'e 4302'
error(['CONFIG The Loss application cannot run,'...
' because it needs one of the modules'...
' is already being used.']);
otherwise
end
end