@@ -146,18 +146,15 @@ Eurotherms are used to control temperature Orange Cryostat, CCR and Furnace devi
146
146
1. See also [#3784](https://github.com/ISISComputingGroup/IBEX/issues/3784)
147
147
148
148
##### Note: Hitachi #####
149
-
Hitachi L-7100 HPLC pump (no longer supported at ISIS). ~Can't find L-7100 on [Hitachi web-site](http://www.hitachi-hightech.com/global/about/corporate/group/hhs (may be obsolete). Check existing SECI VI for logic and manual.~
150
-
* Update (27-06-2019): Hitachi pumps are being retired. No longer any need to support them. See [#3780](https://github.com/ISISComputingGroup/IBEX/issues/3780)
151
-
* SANS2D will not be using Hitachi pumps in future. They will use the [Knauer](#note-knauer-hplc) or Jasco HPLC pumps.
149
+
Hitachi pumps are being retired. No longer any need to support them. SANS2D will not be using Hitachi pumps in future. They will use the [Knauer](#note-knauer-hplc) or Jasco HPLC pumps.
152
150
153
151
##### Note: Knauer #####
154
152
1. Knauer K-6 Electric Valve Drive. [Knauer web site](https://www.knauer.net)<br>
155
-
* K-6 model appears to have been superseded. Check existing SECI VI for logic and manual.
153
+
* K-6 model appears to have been superseded.
156
154
1. See also [#3781](https://github.com/ISISComputingGroup/IBEX/issues/3781)
157
155
158
156
##### Note: Knauer HPLC #####
159
157
Knauer HPLC 1050 [Knauer HPLC 1050 is discontinued.](https://www.knauer.net/en/discontinued-smartline-pump-1050-successor-azura-p-61l/p14161).
160
-
1. Check existing SECI VI for logic and manual.
161
158
1. Support for Knauer HPLC 1050 is implemented via the following tickets:
@@ -236,7 +233,7 @@ SANS2D has two baffle trolleys (in the vacuum tank). One for the front baffle,
236
233
##### Note: Beamstops #####
237
234
SANS2D has beamstops on the front and rear detectors
238
235
1. Front Detector:
239
-
* has a single beamstop, mounted on the detector trolley. Rotates in the plane of the detector array. The beamstop must not be moved when the detector is rotated (in SECI this requirement is implemented as a software inhibit).
236
+
* has a single beamstop, mounted on the detector trolley. Rotates in the plane of the detector array. The beamstop must not be moved when the detector is rotated (in SECI this requirement was implemented as a software inhibit).
240
237
* has a strip beamstop, mounted on the detector trolley. Moves linearly across the face of the detector array. Primarily used for commissioning purposes (usually "parked" out of the way).
241
238
1. Rear Detector:
242
239
* has three beamstops, mounted on a pillar attached to the detector trolley. Pillar moves linearly across the face of the detector array. The individual beam stops move independently up/down on the pillar.
@@ -267,21 +264,11 @@ SANS2D has the following devices under motion control:
267
264
Specialist requirements:
268
265
1. There is a requirement within SANS2D to inhibit the movement of the detectors when they are powered (the CAEN HV is ON)
269
266
270
-
## SANS2D SECI Configs ##
271
-
Document information about SANS2D SECI configs here.
272
-
273
-
Configuration Name | Sub-Configurations | Last Accessed | Required |
Similarly, Document information about SANS2D SECI Genie scripts here.
280
267
281
268
## SANS2D Baldor Drives ##
282
269
SANS2D has 5 guides on the front end. Each of these has a significant weight - ~3 tonnes. The motors used to lift the guides are servo motors. Galil controllers cannot provide sufficient power to motors capable of lifting such weights. As a result, some Baldor drives were introduced to provide the power required to lift these guides. They are digital motion controllers that can communicate with a control system and have settings in each controller to drive the motor.
283
270
284
-
However, ISIS uses them as pseudo dumb power amplifiers only. A Galil drives the axes like a servo - it provides an analogue signal to the Baldor to ask it to drive, but no core power. The Baldor drives the axes. Therefore, from the point of view of SECI or IBEX, the Baldor drives appear as Galil axes.
271
+
However, ISIS uses them as pseudo dumb power amplifiers only. A Galil drives the axes like a servo - it provides an analogue signal to the Baldor to ask it to drive, but no core power. The Baldor drives the axes. Therefore, from the point of view of IBEX, the Baldor drives appear as Galil axes.
285
272
286
273
The Baldor drives do need to be set up – but this is done independently. Historically, this entire arrangement caused a lot of problems during commissioning. There is a bespoke homing routine in use with this setup (the need for such a thing has been questioned but, for now, a bespoke homing routine continues in use).